beautypg.com

Calibration mode – YSI 910 COD Colorimeter User Manual

Page 10

background image

10

Calibration Mode

4

user calibration (cal)

After confirming the selection with the [MODE] key the instrument will
show Cal/lr.

Scroll through ranges using the [MODE] key.
Place the blank

in the adapter making sure that the marks are aligned.

Press the [ZERO/TEST] key.
The “Range“ symbol flashes for approx. 8 seconds.
The display shows the following in alternating mode:
Perform measurement with a standard of known concentration as described.

Press the [ZERO/TEST] key.
The “Range“ symbol flashes for approx. 3 seconds.
the result is shown in the display, alternating with Cal.
If the reading corresponds with the value of the calibration standard (within the

specified tolerance), exit calibration mode by pressing the [ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Pressing the [ZERO/TEST] key once decreases the displayed value by
1 digit.
Press the corresponding key until the reading equals the value of the calibration
standard.
By pressing the [ON/OFF] key, the new correction factor is calculated and stored

in the user calibration software.
Confirmation of calibration (3 seconds).

note:

The instrument calibration in the range MR is automatically taken also for

Hr.

user calibration (Display in calibration mode)
factory calibration (Display in calibration mode)

CAL

cAL

Time Cal

Store Date

4

Cal

CAL

LR

0.0.0

CAL

Zero

Test

Zero

Test

RANGE

RANGE

CAL

RESULT + x

CAL

Mode

On

Off

Zero

Test

: :

RESULT