technical Data

instrument:

double

wavelength, automatic wavelength selection,

direct reading colorimeter

light source:

leDs,

interference filters (IF) and photosensor in transparent

cell chamber. Wavelength specifications of the IF:

430 nm = 5 nm

610 nm = 6 nm

Wavelength accuracy:

± 1 nm

photometric accuracy*:

3% FS (T = 20° C – 25° C)

photometric resolution: 0.01 a

power supply:

4 batteries (AAA/LR 03)

lifetime: approx. 17 hours or 5000 tests

auto-OFF:

instrument automatically turns off

10 minutes after last keypress

Display:

backlit LCD (on keypress)

storage:

internal ring memory for 16 data sets

serial interface:

IR interface for data transfer

time:

real time clock und date

Calibration:

user and factory calibration

resetting to factory calibration possible

Dimensions:

155 x 75 x 35 mm (lxWxH)

Weight:

approx. 260 g (incl. batteries)

ambient conditions:

temperature: 5–40°C

rel. humidity: 30 – 90 % (non-condensing)

Waterproof:

floating; IP 68 (1 hour at 0.1 meter)

Ce:

Certificate for Declaration of CE-Conformity


*measured with standard solutions

To ensure maximum accuracy of test results, always use the reagent systems