beautypg.com

Methods – YSI 900 Chlorine Colorimeter User Manual

Page 8

background image

8

Methods

!

Zero

Test

Zero

Test

CL 6

Tablet Powder

CL 6

Tablet Powder

CL 6

Tablet Powder

0.0.0

Tablet Powder

RESULT

Tablet Powder

RESULT

Tablet Powder

Chlorine with tablet: 0.01 – 6.0 mg/l

a) Free Chlorine
Fill a clean vial (24 mm Ø) with

10 ml of the water sample and perform zero

calibration (see “Operation”).
Remove the vial from the sample chamber and

empty it, leaving a few drops

remaining in the vial.
Add

one DpD no. 1 tablet straight from the foil to the water sample and crush the

tablet using a clean stirring rod.
Add the water sample to the 10 ml mark.
Close the vial tightly with the cap and swirl gently several times until the tablet is

dissolved.
Place the vial in the sample chamber making sure that the marks are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.

b) total Chlorine
Add

one DpD no. 3 tablet straight from the foil to the same water sample and crush

the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until the tablet is

dissolved.
Place the vial in the sample chamber making sure that the marks are aligned.
Wait for a reaction period of 2 minutes.

(Countdown can be activated, see page 7)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.

c) Combined Chlorine
combined Chlorine = total Chlorine – free Chlorine

tolerances:
0 – 1 mg/l: ± 0.05 mg/l

> 1 – 2 mg/l: ± 0.10 mg/l

> 2 – 3 mg/l: ± 0.20 mg/l

> 3 – 4 mg/l: ± 0.30 mg/l

> 4 – 6 mg/l: ± 0.40 mg/l

!

Zero

Test

Zero

Test

CL 6

Tablet Powder

CL 6

Tablet Powder

CL 6

Tablet Powder

0.0.0

Tablet Powder

RESULT

Tablet Powder

RESULT

Tablet Powder