beautypg.com

YSI 900 Chlorine Colorimeter User Manual

Page 13

background image

13

ENGLISH

notes:

1.

Vial cleaning:

As many household cleaners (e.g. dishwasher detergent) contain reducing substances, the

subsequent determination of Chlorine may show lower results. To avoid any measurement errors,

only use glassware free of Chlorine demand.

preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for one hour,

then rinse all glassware thoroughly with deionized water.

2. For individual testing of free and total Chlorine, the use of different sets of glassware is

recommended (EN ISO 7393-2, 5.3)

3. Do not use the same sample vial for free and total chlorine without thoroughly rinsing the vial

between the two different tests.

4.

preparing the sample:

When preparing the sample, the lost of Chlorine, e.g. by pipetting or shaking, must be avoided.

The analysis must take place immediately after taking the sample.

5. The DPD color development is carried out at a pH value of 6.2 to 6.5. The reagents therefore

contain a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted

between pH 6 and pH 7 before the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l

Sodium hydroxide).

6.

exceeding the measuring range:

Concentrations above 2 mg/l Chlorine can lead to results showing 0 mg/l. In this case, the water

sample must be diluted with water free of Chlorine and the measurement repeated.

7. Oxidizing agents such as Bromine, Ozone etc. interfere as they react in the same way as

Chlorine.