Menu options – YSI 900 Chlorine Colorimeter User Manual
Page 16
16
Menu Options
Menu selections
Press the [MODE] key and hold.
Turn the unit on using the [ON/OFF] key.
Allow the 3 decimal points to be displayed before releasing the [MODE] key.
The [!] key allows for selection of the following menu points:
1
diS: recall stored data
1
Prt: printing stored data
2
3
setting the date and time
Cal user calibration
1
dis – recall of stored data
After confirming the selection with the [MODE] key the colorimeter shows the last
16 data sets in the following format (automatically proceeds every 3 seconds until
result is displayed):
Number
n xx (xx: 16...1)
Year
YYYY (e.g. 2010)
Date
mm.dd (monthmonth:dayday)
Time
hh:mm (hourhour:minuteminute)
Test
Method
Result
x.xx
The [ZERO/TEST] key repeats the current data set.
The [MODE] key scrolls through all stored data sets.
Quit the menu by pressing [!] key.
1
prt – transmitting stored data (to printer or pC)
Note: To print data, or to transmit to a PC, the optional
900 series Data Hub (Inf-
rared Interface Module) is required.
The
900 series Data Hub Module and the connected printer/PC must be ready.
Press the [MODE] key to start the transmitting, the instrument displays “PrtG”
(Printing) for approx. 1 second followed by the number of the first data set and its
transmission. All data sets will be transmitted one after the other. After finishing the
instrument returns to test mode.
The print job can be cancelled by pressing the [On/Off] key. The instrument turns off.
On
Off
On
Off
Mode
!
!
Mode
Mode
Zero
Test
PrtG
Store Date
Print Time
2
1
Cal
diS
1
Store Date
Print Time
Prt
1
Store Date
Print Time
1
1
2