What to do if...

80

Buffer solutions:

– Incorrect buffer solutions

– Change calibration

procedure

– Buffer solutions too old

– Only use once.

Note the shelf life

– Buffer solutions depleted

– Change solutions

Cause

Remedy

pH electrode:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

Sample:

– pH value not stable

– Measure with air

excluded if necessary

– Temperature not stable

– Adjust temperature if

necessary

Electrode + sample:

– Conductivity too low

– Use suitable electrode

– Temperature too high

– Use suitable electrode

– Organic liquids

– Use suitable electrode

Cause

Remedy

– Batteries almost depleted

– Replace batteries (see

section 5.1 M

AINTENANCE

)

No stable

measured value

LoBat