Operation

72

5 Thoroughly rinse the electrode with distilled water.

6 If necessary, set the temperature of the second buffer

solution by pressing <

> <

>.

7 Submerse the electrode in the second buffer solution.

8 Press the key.

AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized, AR disap-
pears.
The sensor symbol shows the electrode evaluation
after the two-point calibration.
The value of the slope (mV/pH) appears on the dis-
play.

9 Press the key.

The value of the asymmetry (mV) appears on the dis-
play.

10 To return to the measuring mode: Press the

key.