2 measuring – Xylem CG 842 User Manual

Page 21

background image

Operation

64

4.2

Measuring

Preparatory

activities

Perform the following activities when you want to measure:

Note
Incorrect calibration of the pH electrode will result in incor-
rect measured values. Therefore, regularly perform calibra-
tion before measuring.

Temperature probe

Measurements can be performed with and without a temper-
ature probe. A connected temperature probe is indicated by
TP on the display.

Note
The pH meter automatically recognizes the type of the tem-
perature probe used. As a result, you can connect elec-
trodes with the NTC 30 or Pt1000.

The temperature measurement is absolutely essential for a
reproducible pH measurement. If the measurement is made
without a temperature probe, proceed as follows:

Note
When calibrating without a temperature probe, set up the
current temperature of the respective buffer solution manu-
ally by pressing the <

> <

> keys.

1 Connect the pH electrode to the instrument.

2 Adjust the temperature of the buffer or test solutions

or measure the current temperature if the measure-
ment is made without a temperature probe.

3 Calibrate or check the instrument with the pH elec-

trode according to section 4.3.

4 Select the measuring mode by pressing .

1 Determine the current temperature using a thermo-

meter.

2 Set up the temperature by pressing <

> <

>.