beautypg.com

2 resetting meter settings, Resetting meter settings, Ti n i – Xylem Lab 970 User Manual

Page 50

background image

Operation

Lab 970

50

ba75558e03

11/2009

4.9.2

Resetting meter settings

This function resets to the default condition meter settings. The
relevant values are given on the following pages:

The following settings are also reset to the default condition:

Resetting the meter

settings

System settings

page 44

Measurement settings

page 46

Setting

Default settings

Measured variable

ӕ

mS/cm or µS/cm

Adjusted cell constant

0,650 1/cm

Temperature compensation

nLF

Temperature coefficient of the
linear temperature
compensation

2.000 %/K

TDS factor

1.00

1

Switch on the meter with .
The display test appears briefly on the display.

2

During the display test, press to open the menu for
the reset of the meter settings.
Init is displayed.

3

Press <

▲><▼> to display no or YES.

YES: Reset the meter settings.
no: Retain the meter settings.

o

n

t

i

n

I