1 options for data transmission, 3 rs232 interface, Options for data transmission – Xylem Lab 970 User Manual

Page 40: Rs232 interface

background image

Operation

Lab 970

40

ba75558e03

11/2009

4.7.1

Options for data transmission

Via the USB interface you can transmit data to a PC. Via the RS 232
interface, you can transmit data to a PC or an external printer.

The following table shows which data are transmitted to the interface in
which way:

4.7.2

Automatically downloading measurement datasets at
intervals

In order to automatically download to the interface measured values at
certain time intervals, set the download interval (Int.2).

Setting the download

interval

The default setting for the download interval (Int.2) is OFF.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):

4.7.3

RS232 interface

Data

Control

Operation / description

Current measured
values

manual

z With .

automatic, at
intervals

z With

_

>. Then you

can set the transmission
interval (Int.2) (see
page 40).

automatic

z After each measurement

with stability control.

calibration records

manual

z During the display

indication with

_

>

(see page 38).

automatic

z On completion of a

calibration procedure.

1

Press

_

> to open the setting of the Int.2 interval.

2

If necessary, set an interval with <

▲><▼>.

3

Close the setting with .
The download to the interface takes place at the specified
interval.

1

Connect the interface to the PC or printer via the cable Z390
(PC) or Z393 (ext. printer).