beautypg.com

Xylem Lab 970 User Manual

Page 27

background image

Lab 970

Operation

27

ba75558e03

11/2009

Note
If the error message E3 appears, refer to

CHAPTER

6 W

HAT

TO

DO

IF

...

(

PAGE

55).

Stability control

During calibration, the stability control is automatically activated.

Note
This method of automatically determining the cell constant by
calibration in the 0.01 mol/l KCL control standard can only be used for
measuring cells with cell constants in the ranges
0.450 ... 0.500 cm

-1

, 0.585 ... 0.715 cm

-1

or 0.800 ... 1.200 cm

-1

.

2

Press or

_

> to confirm the selection of CAL

CELL.

3

Immerse the measuring cell in the control standard solution,
0.01 mol/l KCI.

4

Start the calibration with .
The determination of the cell constant with stability control
starts.
The

display indicator flashes until there is a stable signal.

The determined cell constant is displayed for 10 seconds. The
meter stores the cell constant automatically.
After this the meter switches to the measuring mode.