Xylem Lab 850, starting from serial number 09470315 User Manual
Page 56
What to do if...
Lab 850
56
ba75890e01
11/2009
No stable measured
value
CalClock flashes
Display,
LoBat
– Incorrect buffer solutions
– Change calibration
procedure
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change solutions
Cause
Remedy
Cause
Remedy
pH electrode:
– Junction contaminated
– Clean junction
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Adjust temperature if
necessary
Electrode + test sample:
– Conductivity too low
– Use a suitable electrode
– Temperature too high
– Organic liquids
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see