2 resetting all meter settings, Resetting all meter settings, Ti n i – Xylem Lab 850, starting from serial number 09470315 User Manual
Page 52
![background image](/manuals/310551/52/background.png)
Operation
Lab 850
52
ba75890e01
11/2009
4.8.2
Resetting all meter settings
This function resets to the default condition all meter settings including
the calibration values. The corresponding values are given on the
following pages:
Resetting the meter
settings
Note
The measuring system is not calibrated after a reset. Before measuring
recalibrate the meter.
Calibration values
System settings
Measurement settings
1
Switch on the meter with
The display test appears briefly on the display.
2
During the display test, press
the reset of the meter settings.
Init is displayed.
3
Press <
▲><▼> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.
4
Confirm with
The menu is finished.
The meter switches to the measuring mode.
o
n
t
i
n
I