2 Quality Control Certification
Date:
Customer: Order No.:
Pass
Model
GPR-2500S Ambient Oxygen Monitor
Sensor
( ) GPR-11-32-4 Oxygen Sensor
( ) XLT-11-24-4 Oxygen Sensor
( ) Other _____________________
Serial Nos.:
Monitor: Sensor:
Accessories:
Owner’s Manual
( ) A-2344 Calibration Flow Adapter (GPR-2500S)
( ) Remote sensor requires: A-2070 Bracket Mount, A-2781 Nut Retaining
Configuration:
A-1151-E-L2 PCB Assembly Main / Display
Software rev:
Ranges: 0-1%, 0-5%, 0-10%, 0-25%
Power: 12-36V DC two wire loop power
NEMA 4 rated wall mount enclosure
Barometric pressure compensation
Test:
Default zero (without sensor)
Default span @ 40uA
Analog signal output 4-20mA full scale
Calibrates with adequate span adjustment within 10-50% FS
Baseline drift on zero gas <
± 2% FS over 24 hour period
Noise level <
± 1.0% FS
Span adjustment within 10-50% FS
Final:
Overall inspection for physical defects
Options:
Notes: