2 quality control certification – Analytical Industries AII-4001 Helium Oxygen Trimix Analyzer User Manual
Page 3

2 Quality Control Certification
3
4
Date: ________ Customer _____________________________ Order _____________
Model:
AII-4001 Helium Oxygen Analyzer
S/N _____________
Sensors:
AII-11-75D Oxygen Sensor
S/N _____________
AII-41-100 Helium Sensor
S/N _____________
AII-E3 Environmental Sensor Module
S/N _____________
Configuration: A-1185 PCB Assembly, Main Software Ver ………... _____________
Accessories:
A-3582 Flow Manifold ……………………………………………………….. ______
A-3609 Adapter, Dome to 1/8” Tube …...…………………………….. ______
TUBE-1018 Tubing 1/8” x 3 ft …………………………………………….. ______
PWRS-1003-1 Charger 9VDC 110/220 VAC Adapter ……………... ______
Instruction Manual ………………………………………………………………. ______
Testing:
Mechanical: all screws, sensors and connectors secured …………. ______
Labels: 3 labels applied to case …………………………………………….. ______
LCD display 3-1/2 digits ……………………………………………………….. ______
Backlight: on for 3 seconds after power-up ……………….…………...
_____
on for 15 seconds after key press ………………………….. ______
Battery icon: “LOW BATTERY” displays in reverse background
in lower right hand corner when the battery icon
displays < 30% charge ………………………..…………. ______
displays solid black when fully charged ….……….. ______
Charging battery: “CHARGING” displays in lower right hand
corner when the charge plug is connected
to the analyzer and the adapter plugged into
a 110/220 VAC outlet ……………………...……..…
______
Test (cont’d)
Temperature: within 2⁰F of EXTECH RH300 Digital Psychrometer
at ambient conditions ……………………………………….. ______
Relative Humidity: within 3% of EXTECH RH300 Digital
Psychrometer at ambient conditions ……….….. ______
Barometric Pressure: within 1 KPa of Accu Cal Plus Digital
Precision Test Gauge ……………………………... ______
Calibration O2 Sensor: +10-30% FS in ambient air ………………..… _______
>98% in 100% oxygen ……………………….. ______
Zero He Sensor: +1% FS in 100% oxygen ………………………………. ______
>98% in 100% helium span gas …..……………….. ______
O2 Sensor: within 0.5% of ambient air calibration value …………... ______
Overall Appearance: clean, no physical defects ………..………………. ______
Part Number & Description _Qty_
Options:
PWRS-1019 Adapter, 9VDC Cigarette Lighter …………………………… ______
A-3673 Adapter, BC with Flow Resistor to 1/8” Tube ………...……. ______
A-3677 Adapter, DIN to 1/8” Tube ………………………………………... ______
A-3678 Adapter, A-Yoke to 1/8” Tube ……………………………………. ______
Notes:
Delivery: