Symptom possible cause recommended action – Analytical Industries GPR-1100 ATEX Portable Trace PPM Oxygen Analyzer User Manual
Page 30

 
 
 
30
Symptom
Possible Cause
Recommended Action
Erratic O
2
reading
or
Negative O
2
reading
or
No O
2
reading
possibly
accompanied by
electrolyte 
leakage 
Pressurizing the sensor by
flowing gas to the sensor with:
the vent restricted or 
SHUT OFF valve closed 
and
suddenly removing the
restriction draws a vacuum on
the sensor
 or 
partially opening the valves 
upstream of the analyzer when
using a pump downstream of
the analyzer to draw sample
from a process at atmospheric
pressure or a slight vacuum 
 
A pressurized sensor may not
leak but still produce negative
readings.
Placing a vacuum on the sensor 
in excess 4” of water column is 
strongly discouraged. The front
sensing membrane is .000625
thick, heat sealed to the sensor
body and subject to tearing
when vacuum is suddenly 
applied. 
A premature adjustment of the
ZERO OFFSET potentiometer is
a common problem 
 
Zero the analyzer. If not successful replace
the sensor
 
Avoid drawing a vacuum on the sensor 
 
 
 
 
 
 
 
 
From MAIN MENU select DEFAULT ZERO
 
 
 
 
 
 
 