Symptom

Possible Cause

Recommended Action

Erratic O

2

reading

or

Negative O

2

reading

or

No O

2

reading

possibly

accompanied by

electrolyte
leakage

Pressurizing the sensor by

flowing gas to the sensor with:

the vent restricted or
SHUT OFF valve closed

and

suddenly removing the

restriction draws a vacuum on

the sensor

or
partially opening the valves

upstream of the analyzer when

using a pump downstream of

the analyzer to draw sample

from a process at atmospheric

pressure or a slight vacuum

A pressurized sensor may not

leak but still produce negative

readings.

Placing a vacuum on the sensor
in excess 4” of water column is

strongly discouraged. The front

sensing membrane is .000625

thick, heat sealed to the sensor

body and subject to tearing

when vacuum is suddenly
applied.

A premature adjustment of the

ZERO OFFSET potentiometer is

a common problem

Zero the analyzer. If not successful replace

the sensor


Avoid drawing a vacuum on the sensor





From MAIN MENU select DEFAULT ZERO