Specular Reflectance Standards Operation Instructions

2

STAN-00000-000-01-201106

3. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark

Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This
command merely stores a dark spectrum in memory. You must select File | Save | Dark from the
menu to permanently save the spectrum to disk.

You must take a dark spectrum before measuring transmission.

4. Put the sample is in place and ensure that the light path is clear. Then, take a reflection

measurement by clicking on the Transmission Mode icon on the toolbar or selecting Spectrum |
Transmission Mode
from the menu bar (since the mathematical calculations used to calculate
transmission and reflection are identical). To save the spectrum to disk, click the Save icon on the
toolbar or select File | Save | Processed from the menu bar.

Operating Using SpectraSuite Software

Procedure

1. Place SpectraSuite in Scope mode by clicking the Scope (

) icon in the Experiment mode

toolbar or selecting Processing | Processing Mode | Scope from the menu.

2. Ensure that the entire signal is on scale. The intensity of the reference signal’s peak differs

depending on the device being used. If necessary, adjust the integration time until the intensity is
appropriate for your device. Ensure that the entire signal is on scale.

3. Select File | New | Reflection Measurement. Use the Reflection Measurement wizard to take a

reference spectrum (with the reference standard) and a dark spectrum. You must take a reference
and a dark spectrum before measuring reflection.

4. Put the sample in place and ensure that the light path is clear. Then, take a reflection

measurement.

5. If you received calibration file with your reference standard, click the Non-unity Correction icon

(

) on the toolbar. The Select Reference File screen appears.

6. Browse to the location where you stored your reflectivity standard files. Then click the Apply

button.

The reference information appears on the Select Reference File screen.

Specifications

Specification

STAN-SSH

STAN-SSH-NIST

STAN-SSL

Substrate dimensions

31.75 mm outer diameter x 6.35 mm height

Housing dimensions

38 mm outer diameter x 19 mm height

Weight

40 g

Reflectance material

Front surface-protected
aluminum mirror on fused
silica substrate

Front surface-protected
aluminum mirror on
fused silica substrate

Schott ND9 glass

Reflectivity

200 – 1000 nm: ~87 – 93%
1000 – 2050 nm: ~93 – 98%

200 – 950 nm: ~5%
950 – 2500 nm: ~4%