Specifications, Operating using spectrasuite software – Ocean Optics Specular Reflectance Standards User Manual
Page 2
Specular Reflectance Standards Operation Instructions
2
STAN-00000-000-01-201106
3. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark
Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This
command merely stores a dark spectrum in memory. You must select File | Save | Dark from the
menu to permanently save the spectrum to disk.
You must take a dark spectrum before measuring transmission.
4. Put the sample is in place and ensure that the light path is clear. Then, take a reflection
measurement by clicking on the Transmission Mode icon on the toolbar or selecting Spectrum |
Transmission Mode from the menu bar (since the mathematical calculations used to calculate
transmission and reflection are identical). To save the spectrum to disk, click the Save icon on the
toolbar or select File | Save | Processed from the menu bar.
Operating Using SpectraSuite Software
► Procedure
1. Place SpectraSuite in Scope mode by clicking the Scope (
) icon in the Experiment mode
toolbar or selecting Processing | Processing Mode | Scope from the menu.
2. Ensure that the entire signal is on scale. The intensity of the reference signal’s peak differs
depending on the device being used. If necessary, adjust the integration time until the intensity is
appropriate for your device. Ensure that the entire signal is on scale.
3. Select File | New | Reflection Measurement. Use the Reflection Measurement wizard to take a
reference spectrum (with the reference standard) and a dark spectrum. You must take a reference
and a dark spectrum before measuring reflection.
4. Put the sample in place and ensure that the light path is clear. Then, take a reflection
measurement.
5. If you received calibration file with your reference standard, click the Non-unity Correction icon
(
) on the toolbar. The Select Reference File screen appears.
6. Browse to the location where you stored your reflectivity standard files. Then click the Apply
button.
The reference information appears on the Select Reference File screen.
Specifications
Specification
STAN-SSH
STAN-SSH-NIST
STAN-SSL
Substrate dimensions
31.75 mm outer diameter x 6.35 mm height
Housing dimensions
38 mm outer diameter x 19 mm height
Weight
40 g
Reflectance material
Front surface-protected
aluminum mirror on fused
silica substrate
Front surface-protected
aluminum mirror on
fused silica substrate
Schott ND9 glass
Reflectivity
200 – 1000 nm: ~87 – 93%
1000 – 2050 nm: ~93 – 98%
200 – 950 nm: ~5%
950 – 2500 nm: ~4%