Ocean Optics FluoroVette User Manual
Ocean Optics Equipment
FluoroVette
TM
Product Specifications
Materials:
Biocompatible Black Delrin body bonded with a
Non-leaching solvent acrylic adhesive
to a 50 micron UV transparent film.
Connectors:
Hose barb connector (CFV-PUMP) is Polysulfone seated in acrylic
Pipette interface (CFV-PIP) is acrylic
Contained volume, 50 uL
Pathlength 0.75 mm
Dead Volume in channels: 1 uL for the inlet channel, 1 uL for outlet
channel
FluoroVette Overall dimensions: 4.7 mm tall, 4.3 mm wide, and 1 mm thick.
Adaptor dimensions: 15 mm from base to the center of light path, 1 cm x 1 cm
outer dimension. Insert retaining ring with or without a lens to reduce light scatter
NOTE: The adaptor has a separate retaining ring to allow insertion of a plano-
convex lens. Suggested lens; Edmund Optics: www.edmundoptics.com
LENS PCX 5 X 4.6 MGF2
COM
Dia.
5.0
E.F.L.
4.6
Type
PCX
Comment
LAKN14
Part #
NT32-403
Price
$19.00
Solvent compatibility:
Ethanol ( up to 99.5%)
Isopropyl alcohol ( up to 91% in water)
Aqueous buffers with or without detergents
DO NOT FILL with Acetone or Aliphatic solvents such as Hexane. Wipe
the outside with lens tissue to avoid scratching the Optical Film.
Directions for use:
The FluoroVette has an inlet side, which allows the flowcell to fill from the bottom to
minimize air bubble formation. The inlet side is shown on the left, the outlet side is
shown on the right in Figure 1.
NOTE Insert the FluoroVette
TM
into the adaptor with the inlet side facing
the light source of the fluorometer. Two notches at the top of the adaptor
indicate the light path as shown in Figure 2
Figure 1
Figure 2
FluoroVette
TM
with Hose Barb Connections
Shown below with the Adaptor assembly.
Inlet
Outlet
Side
Side
Light Source
FluoroVette
TM
with Pipette Interface
ALine, Inc. PO Box 1068, Redondo Beach, CA 90278
830 Douglas Ave., Dunedin, FL 34698
Customer
support:
877-707-8575