beautypg.com

Ocean Optics SAD500 Communications and Control User Manual

Page 8

background image

8

Pixel Mode

Description: Specifies which pixels are transmitted or stored in slow memory. While all pixels are acquired
on every scan, this parameter determines which pixels will be transmitted or stored.

Command Syntax:

P{

DATA WORD

}

Value:

Description
0 = all 2048 pixels
1 = every n

th

pixel with no averaging

2 = every n

th

pixel with averaging (effective boxcar

width of n)

3 = pixel x through y every n pixels
4 = up to 81 randomly selected pixels between 0 and

2047 (denoted p1, p2, … p81)

256 = all 2048 pixels, compressed
257 = every nth pixel with no averaging, compressed
258 = every nth pixel with averaging, compressed
259 = pixel x through y every n, compressed
260 = up to 81 selected pixels, compressed
512 = all 2048 pixels, with CDS
513 = every nth pixel with no averaging, with CDS
514 = every nth pixel with averaging, with CDS
515 = pixel x through y every n, with CDS
516 = up to 81 selected pixels, with CDS
768 = all 2048 pixels, compressed, with CDS
769 = every nth pixel with no averaging, compressed,

with CDS

770 = every nth pixel with averaging, compressed, with

CDS

771 = pixel x through y every n, compressed, with CDS
772 = up to 81 selected pixels, compressed, with CDS

Example
P 0

(spaces for clarity only)

P 1 n
P 2 n

P 3 x y n
P 4 n p1 p2 p3…p81

Response:

ACK or NAK

Default value:

0

All pixel modes involving correlated double sampling can only be used with the S1024DW
spectrometer.

Since most applications only require a subset of the spectrum, using the Pixel Mode command can
greatly reduce the amount of time required to transmit a spectrum while still providing all of the
desired data.

Initialize

Description: Sets all operating parameters to their default value.

Command Syntax:

Q

Response:

ACK or NAK

This command resets operating parameters to the parameters that were read from flash memory at
power-up.