3B Scientific Digital Counter with Interface (230 V, 50__60 Hz) User Manual
Page 5

5
every time the push-button ‘START/STOP’ is 
pressed. 
LED ‘Input A’ is lit when the input (the gate) is 
open. 
Selection of trigger polarity is possible, if pendu-
lum is selected. 
The frequency is displayed in Hz or kHz. 
5.2 Period measurement
•
Select frequency measuring with the push-
button ‘FUNCTION’ and hereafter the wan-
ted measuring function with the push-button 
‘SELECT’
.
•
Press the button ‘START/STOP’ to start the 
frequency measurement. 
If the measuring function ‘Single measuring’ or 
‘Pendulum Single ‘ (pendulum), has been se-
lected, a new period measurement is made, 
every time the push-button ‘START/STOP’ is 
pressed. 
Selection of trigger polarity is possible, if pendu-
lum is selected. 
LED ‘Input A’ is lit when the input (the gate) is 
open. 
The period time is displayed in s or ms. 
5.3 Impulse counting
•
Select frequency measuring with the push-
button ‘FUNCTION’ and hereafter the wan-
ted measuring function with the push-button 
‘SELECT’
.
At all counting functions, the ‘Input A’-LED is lit 
when counting. 
If ‘Continuous’ has been selected, the counter is 
started when ‘START/STOP’ is pressed. The 
counter can be stopped and restarted (without 
resetting the counter) by pressing the 
‘START/STOP’
.
If the counter has to be reset (count set to zero): 
push ‘START/STOP’ to stop the counting, and 
thereafter push ‘FUNCTION’ to reset and pre-
pare the counter for a new counting. 
If count time has been selected to 1, 6, 10 or 60 
seconds, a new counting time is started by 
pressing ‘START/STOP’. The counter can not 
be restarted if a count is in progress. 
If ‘Count. 10s-Hold 5s’ has been selected, the 
counter is started when ‘START/STOP’ is pres-
sed. Counting is done in segments of 10 sec-
onds with a 5 seconds’ interval pause. The 
counter is reset before each 10 seconds’ count-
ing. The function is stopped when ‘FUNCTION’ 
is pressed, and the counter is then reset and 
prepared for a new measurement. 
If ‘10s Continuous’ has been selected, the 
counter is started when ‘START/STOP’ is pres-
sed. The counting sequence is continuous, with 
10 seconds’ counting followed by a 5 seconds’ 
displaylock (for reading the result). The counter 
is reset (and restarted) every 10 seconds, even 
if the display is locked (the counter is still count-
ing). The function is stopped when ‘FUNCTION’ 
is pressed, and the counter is reset and pre-
pared for a new measurement. 
5.4 Start A - Stop B
•
The function is selected with the push-button 
‘FUNCTION’
, and started when
‘START/STOP’
is pressed.
The time measurement starts when an impulse 
occurs at input A, and stops when an impulse 
occurs at input B. The time is displayed in s or 
ms. 
The positive- or negative-going triggering is 
setup by pressing ‘SELECT’ and ‘FUNCTION’ 
simultaneously. The current state of the setup is 
shown in the display i.e. 
Pressing ‘SELECT’ shifts through the four dif-
ferent combinations. When satisfied, press 
‘START/STOP’
and the setup is stored in the
internal Lithium Battery backup-memory. 
The LEDs ‘Input A’ and ‘Input B’ are lit until an 
impulse has occurred at each input. When the 
measurement is stopped, a new measurement 
is prepared when ‘START/STOP’ is pressed. 
The function is aborted if ‘FUNCTION’ is 
pressed again, and the counter is reset and 
prepared for a new measurement. 
5.5 Pass times A and B
•
The function is selected with the push-button 
‘FUNCTION’
, and started when
‘START/STOP’
is pressed.
The positive- or negative-going triggering is 
setup by pressing ‘SELECT’ and ‘FUNCTION’ 
simultaneously. The current state of the setup is 
shown in the display i.e. 
Pressing ‘SELECT’ shifts through the four dif-
ferent combinations. When satisfied, press 
‘START/STOP’
and the setup is stored in the
internal Lithium Battery backup-memory. 
The LEDs ‘Input A’ and ‘Input B’ are lit. 
The pass times on ‘Input A’ and ‘Input B’ are 
measured independent from each other. During 
passage of A and B, the light emitting diode at 
‘Input A’
or ‘Input B’ is turned off.