3. Contents

1 Critical potentials tube S with He filling
1 Battery unit (battery not included)
1 Shielding
1 Manual


4. Technical data

Gas filling:

Helium

Filament voltage:

U

F

≤ 7 V DC

Anode voltage:

U

A

≤ 60 V

Anode current:

I

A

≤ 10 mA

Collector voltage:

U

R

= 1,5 V

Collector current:

I

R

≤ 200 pA

Glass bulb:

130 mm diam. approx

Length of tube:

260 mm approx.


5. Additionally required

For operating the tube:
1 Tube holder S

1014525

1 Control Unit for Critical Potential Tubes
(115 V or 230 V)

1000633 / 1008506

1 DC-Power Supply, 0–20 V
(115 V or 230 V)

1003311 / 1003312

or
1 Power Supply Unit for Franck-Hertz Experiment
(115 V or 230 V)

1012819 / 1012818

For making measurements:
1 Analogue Oscilloscope 2x30 MHz 1002727
2 HF Patch Cords, BNC/4 mm Plug

1002748

or
1 3B NETlog

TM

(115 V or 230 V)

1000539 / 1000540

1 3B NETlab

TM

1000544

1 Battery AA 1.5 V
1 Set of 15 Safety Experiment Leads 1002843


6. Operation

6.1 Setting up the tube in the tube holder

The tube should not be mounted or removed
unless all power supplies are disconnected.

Press tube gently into the stock of the holder
and push until the pins are fully inserted.

Take note of the unique position of the guide
pin.

6.2 Removing the tube from the tube holder

Allow the tube to cool before dismantling.

To remove the tube, apply pressure on the
guide pin until the pins loosen, then pull out
the tube.


7. Example experiment

Determine the critical potentials of a helium
atom

7.1 General notes
The experiment set-up with the critical potential
tube is highly sensitive to sources of electro-
magnetic interference. (computers, fluorescent
lights).

Select a location for the experiment where
electromagnetic interference can be
avoided.

7.2 Experiment set-up with the control unit

for critical potential tubes

Insert the tube into the tube holder.

Provision of heater voltage V

F

:

Connect the sockets F3 of the tube holder to
the positive terminal of the DC power supply
and F4 to the negative terminal. (see Fig. 1)

Provision of accelerating voltage V

A

:

Connect socket C5 of the tube holder to the
negative terminal of the output V

A

on the

control unit and to the negative terminal of
the DC power supply.

Connect socket A1 to the positive terminal of
the output V

A

on the control unit.

Provision of collector voltage V

R

:

The positive pole of the output V

A

should be

connected to the negative pole of the 1.5-V
battery.

Connect the positive terminal of the 1.5 V
battery to an earth socket on the control unit.

Put the shielding over the tube and slot it into
the groove on the tube holder so that the tube
is completely enclosed by shielding. Then
connect the control unit by means of the
ground socket.

Connect the lead from the collector ring to
the BNC input socket of the control unit.