beautypg.com

3B Scientific Binocular Microscope Model 400 User Manual

Page 5

background image

2

(with specimen protection)

Magnification: 40x – 1500x

Object stage: x-y cross table, 132 x 145 mm

2

, with

object guide and coaxial adjustment knobs per-
pendicular to the object stage, adjustment range
50 x 76 mm

2

Illumination: Adjustable 6 V, 20 W halogen lamp
incorporated into the base, universal 85 to 265 V,
50/60 Hz power supply

Condenser: Abbe condenser N.A.1.25 NA 0.65 with
iris diaphragm , filter holder and blue filter, focus-
sed via rack and pinion drive

Dimensions: 328 x 214 x 394 mm³ approx.

Weight: 6.1 kg approx.


3. Unpacking and assembly

The microscope is packed in a molded styrofoam
container.

Take the container out of the carton remove
the tape and carefully lift the top half off the
container. Be careful not to let the optical i-
tems (objectives and eyepieces) drop down.

To avoid condensation on the optical compo-
nents, leave the microscope in the original pa-
cking to allow it to adjust to room tempera-
ture.

Using both hands (one around the pillar and
one around the base), lift the microscope from
the container and put it on a stable desk.

The objectives will be found within individual
protective vials. Install the objectives into the
microscope nosepiece from the lowest magni-
fication to the highest, in a clockwise direction
from the rear.

Put the binocular head onto the top of the
stand and tighten the head-lock-screw. Insert
the eyepieces into the tube.


4. Operation

4.1 General information

Set the microscope on a level table.

Place the object to be observed in the centre of
the specimen stage and clamp it to the object
guide.

Connect the mains cable to the net and turn on
the switch to get the object illuminated.

Make certain that the specimen is centered
over the opening in the stage.

Adjust the interpupillary distance so that one
circle of light can be seen.

Make the necessary eyepiece dioptre adjust-
ments to suit your eyes.

To obtain a high contrast, adjust the back-
ground illumination by means of the iris dia-
phragm and the variable illumination control.

Rotate the nosepiece until the objective with
the lowest magnification is pointed at the
specimen. There is a definite “click” when each
objective is lined up properly.

NOTE: It is best to begin with the lowest power
objective. This is important to reveal general struc-
tural details with the largest field of view first.
Than you may increase the magnification as nee-
ded to reveal small details.

When 100x (oil) objec-

tive is chosen, objective oil must be dripped onto
the slide.

To determine the magnification at which you are
viewing a specimen, multiply the power of the
eyepiece by the power of the objective.

Adjust the holding brake to give a suitable
degree of tightness in the focusing mechanism.

Adjust the coarse-focusing-knob which moves
the stage up until the specimen is focused. Be
careful that the objective does not make con-
tact with the slide at any time. This may cause
damage to the objective and/or crack your
slide.

Adjust the fine-focusing-knob to get the image
more sharp and more clear.

Colour filters may be inserted into the filter
holder for definition of specimen parts. Swing
the filter holder out and insert colour filters.

Use the knobs of the mechanical stage to move
the slide side-, back- and forwards. The vernier
provides accurate location of the specimen
area.

4.2 Changing the lamp and fuse

4.2.1 Changing the lamp

Turn off the power switch, unplug the mains
plug and let the lamp cool down to avoid being
burnt.

For safety reasons, remove the eyepiece.

To change the lamp lay the microscope on its
back to reach the lid on the bottom side.

Loosen screw C of the lamp socket and push it
outwards so that it is in the position shown in
Fig.1.

Loosen screw A and open the cover.

To remove the halogen lamp, use a cloth or
similar material. Do not touch the bulb with
the bare hand.

Lift out the halogen lamp and replace it with a
new one.