7. Troubleshooting

All errors or failures will be shown as codes. The following table explains
the meaning of codes and solutions.

Code

Cause

Corrective Measure

204

Calculation error

Refer to user menu, repeat the
procedures.

208

Currency for LD
out of range

If cannot be solved after several times of
reboot, meaning the instrument is out of
work, please contact your distributor.

220

Battery too low

Recharge the battery

253 Temperature

too

low

Warm up instrument

255

Receiver signal
too weak,
measurement
time too long

Use target plate or change a good
refection.

256

Received signal
too strong

Target too reflective, use target plate or do
not aim at strong light objective.

Error

Hardware error
and uncertainty
error

Switch on/off the device several times. If
the symbol still appears, then your
instrument is defective. Please contact
your distributor.