Appendix b - technical specifications – Velleman ESR70 User Manual
Page 12

Atlas ESR User Guide
February 2008 – Rev 4
Page 12
Appendix B - Technical Specifications
All values are at 25
°C unless otherwise specified.
Parameter
Min
Typ
Max
Note
Peak test current into S/C
±20mA
±25mA
Peak test voltage across O/C
±2.5V
±3.0V
Capacitance measurement range
1µF
22,000µF
Capacitance accuracy
±4% ±0.2µF
ESR measurement range
0Ω 20Ω 2
ESR resolution for ESR < 2Ω 0.01Ω 0.02Ω
ESR resolution for ESR > 2Ω 0.1Ω
0.2Ω
ESR accuracy for ESR < 2Ω ±1.5%
±0.02Ω
ESR accuracy for ESR > 2Ω ±1.5%
±0.2Ω
Abuse voltage (for C < 10µF) ±275V
3
Abuse voltage (for C > 10µF) ±50V
3
Auto-Discharge voltage limit
±50V
Battery type
MN21/GP23A 12V Alkaline
2
Battery voltage range
8.5V
12V
Battery voltage warning threshold
8.5V
Inactivity power-down period
30 seconds
Dimensions (excluding test leads)
103 x 70 x 20 mm
Operating temperature range
10
°C
40
°C
1
Notes
1.
Subject to acceptable LCD visibility.
2.
Subject to revision.
3.
Maximum abuse voltage rated limitation of internal protection electronics.