Capacitor detection, Resistor detection – Velleman LCR40 User Manual
Page 16

Atlas LCR User Guide
January 2009 – Rev 4
Page 16
Capacitor Detection
The Atlas LCR will tell you that you have a capacitor if the following criteria
are satisfied:
1. If the measured DC resistance is higher than 10MΩ, even if the
measured capacitance is very low (such as open probes).
or
2. If the measured DC resistance is between 100kΩ and 10MΩ and
the measured capacitance is larger than 10pF.
or
3. If the measured DC resistance is between 1kΩ and 100kΩ and the
measured capacitance is larger than 100nF.
Resistor Detection
Measured characteristics that do not satisfy any of the above criteria (for
inductors or capacitors) will be displayed as a resistive element.
See also other documents in the category Velleman Accessories for electrical:
- PS1502A (6 pages)
- VMB6PBN (15 pages)
- VTTEST14 (5 pages)
- VMB3PS (10 pages)
- PSSE60 (18 pages)
- VMB8IR (10 pages)
- PSSE24 (4 pages)
- VMBLCDWB (12 pages)
- VMB1TSW (24 pages)
- VL3288 (5 pages)
- PSSMV24 (23 pages)
- VMB4IRT (16 pages)
- VMB7IN (8 pages)
- VMB1USB (6 pages)
- PSI600B (29 pages)
- VL06LA (4 pages)
- PSSEUSB6A (2 pages)
- PS603 (17 pages)
- VMB1TS (43 pages)
- PSSE23 (4 pages)
- VL1212 (29 pages)
- VL7168 (7 pages)
- PSSE45 (19 pages)
- PS925 (13 pages)
- VMB1BL (12 pages)
- VMB1RS (6 pages)
- PSS1320 (11 pages)
- PSSMV8 (20 pages)
- VMB8PB (16 pages)
- VMB4RF (8 pages)
- VMBRSUSB (8 pages)
- PI300BN (33 pages)
- VMBGPTCx (12 pages)
- PSC1350 (2 pages)
- VTTEST15 (7 pages)
- PSSEUSB4 (2 pages)
- VMB4DC (18 pages)
- VMBGP1x (12 pages)
- PSSMV2 (4 pages)
- VMB8PBU (12 pages)
- PSIC75B (18 pages)
- PS12015 (5 pages)
- VMB4TX (1 page)
- VL6278 (12 pages)
- VMB4PD (35 pages)