Velleman DVM1200 User Manual
Page 12
DVM1200
VELLEMAN
- 12 -
Temperature
Range Resolution
Accuracy
-55C°~0°C
0.1°C
± (5.0% of rdg ± 4°C)
1°C~400°C
± (2.0% of rdg ± 3°C)
401°C~1,000°C
1°C
± 2.0% of rdg
Temperature specifications do not include thermocouple errors
Overload Protection: 250VDC or 250VAC RMS
Capacitance
Range Resolution
Accuracy
60nF
10pF
± (3.0% of rdg + 20 digits)
600nF 100pF
± (3.0% of rdg + 10 digits)
6µF 1nF
60µF 10nF
300µF
100nF
± (5.0% of rdg + 10 digits)
Overload Protection: 250VDC or 250VAC RMS
Current
DC Current
Range Resolution
Accuracy
600µA 0.1µA
± (1.5% of rdg ± 3 digits)
6000µA 1µA
60mA 0.01µA
± (1.5% of rdg ± 5 digits)
600mA 0.1µA
6A 1mA
± (2.0% of rdg ± 5 digits)
10A 10mA
AC Current
Range Resolution
Accuracy
600µA 0.1µA
± (1.8% of rdg ± 5 digits)
6000µA 1µA
60mA 0.01µA
± (2.0% of rdg ± 8 digits)
600mA 0.1µA
6A 1mA
± (3.0% of rdg ± 8 digits)
10A 10mA
1
Frequency Range for ACA: 40Hz~400Hz
2
Response for ACA: avg, calibrated in RMS of sine wave
Overload Protection: F10A/600V fuse for 10A range; F750mA/600V fuse for µA and mA ranges
Max. Input Current: 600mA DC or 600mA AC RMS for µA and mA ranges; 10A DC or 10A AC RMS for 10A ranges
For measurements > 6A, max. 4 minutes ON to measure, 10 minutes OFF; above 10A unspecified
Use this device with original accessories only. Velleman nv cannot be held responsible in the event of
damage or injury resulted from (incorrect) use of this device.
For more info concerning this product, please visit our website www.velleman.eu.
The information in this manual is subject to change without prior notice.