AC Voltage
Range Resolution
Accuracy
600mV
0.1mV
± (3.0% of rdg ± 3 digits) (manual)
6V 1mV
± (1.0% of rdg ± 3 digits)
60V 10mV 
600V 100mV 
700V 
1V
± (1.5% of rdg ± 3 digits)
1
Frequency Range for ACV: 40Hz~400Hz
2
Response for ACV: avg, calibrated in RMS of sine wave
 
Overload protection: 1,000VDC or 700VAC RMS 
Input impedance (nominal): > 10MΩ; AC voltage: > 10MΩ 
Common mode rejection ratio: DC voltage: > 100dB @ DC, 50 or 60Hz; AC voltage: > 60dB @ DC, 50 or 60Hz 
Normal mode rejection ratio: DC voltage: > 45dB @ 50 or 60Hz 
 
Frequency 
 
Range Resolution 
Accuracy
99.99Hz 0.01Hz
± (0.1% of rdg ± 3 digits)
999.9Hz 0.1Hz 
9.999kHz 0.001kHz 
99.99kHz 0.01kHz 
999.9kHz 0.1kHz 
Overload protection: 250VDC or 250VAC RMS 
 
Resistance 
 
Range Resolution 
Accuracy
600.0Ω 0.1Ω
± (0.5% of rdg ± 3 digits)
6.000kΩ
1Ω
± (0.5% of rdg ± 2 digits)
60.00kΩ 10Ω 
600.0kΩ 100Ω 
6.000MΩ 1kΩ 
60.00MΩ 10kΩ 
± (1.5% of rdg ± 3 digits)
Overload protection: 250VDC or 250VAC RMS
 
Diode Test 
 
Range Resolution 
Accuracy
1V 0.001V
1.0%
uncertainty
Forward DC Current: ± 1mA
Reverse DC Voltage: ± 1.5V 
Overload Protection: 250VDC or 250VAC RMS 
Continuity Check 
 
Range Resolution 
Accuracy
600Ω 0.1Ω
Open circuit voltage approx. 0.5V
Continuity beeper: ≤ 50Ω 
Overload Protection: 250VDC or 250VAC RMS 