6 Set up with the display and adjustment module
VEGABAR 86 • Slave sensor for electronic differential pressure
45052-EN-131011
1. Select the menu item "Setup" with [->] and confirm with [OK].
Now select with [->] the menu item "Max. adjustment" and con-
firm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the
requested position with [->].
3. Set the requested percentage value with [+] and save with [OK].
The cursor jumps now to the density value.
4. Enter the max. density value matching the percentage value.
The max. adjustment for density is finished.
Proceed as follows:
1. Select the menu item "Setup" with [->] and confirm with [OK].
Now select with [->] the menu item "Min. adjustment" and confirm
with [OK].
2. Edit the percentage value with [OK] and set the cursor to the
requested position with [->].
3. Set the requested percentage value with [+] and save with [OK].
The cursor jumps now to the height value.
4. Enter the min. height of the interface suitable for the percentage
value.
5. Save settings with [OK] and move with [ESC] and [->] to the max.
adjustment.
The min. adjustment, interface is hence finished.
Proceed as follows:
1. Select the menu item "Setup" with [->] and confirm with [OK].
Now select with [->] the menu item "Max. adjustment" and con-
firm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the
requested position with [->].
3. Set the requested percentage value with [+] and save with [OK].
The cursor jumps now to the height value.
4. Enter the max. height of the interface matching the percentage
value.
The max. adjustment for interface is finished.
Setup - Min. adjustment -
Interface
Setup - Max. adjustment -
Interface