beautypg.com

4 differential pressure measurement, 5 interface measurement – VEGA VEGABAR 86 Save sensor - Operating Instructions User Manual

Page 15

background image

15

4 Mounting

VEGABAR 86 • Slave sensor for electronic differential pressure

45052-EN-131011

4.4 Differential pressure measurement

With the Master/Slave combination, a gauge different measurement

can be realized.

2

3

1

Fig. 7: Measurement setup with a level difference measurement

4.5 Interface measurement

With the Master/Slave combination also an interface measurement

can be realised.
Requirements for a functioning measurement are:

Vessel with changing level

Products with steady densities

Interface always between the measurement points

Total level always above the upper measurement point

Measurement setup