beautypg.com

VEGA VEGAFLEX 83 Profibus PA Polished version User Manual

Page 79

background image

79

11 Supplement

VEGAFLEX 83 • Profibus PA

44223-EN-130910

L

0

-2 mm

(-0.079

")

2 mm

(0.079

")

15 mm

(0.591

"

)

-15 mm

(-0.591

")

0,2 m

(7.874

"

)

0,08 m

(3.15

"

)

0,04 m

(1.575

")

1

Fig. 54: Deviation VEGAFLEX 83 in rod version in water
1 Dead band - no measurement possible in this area

L Probe length

L

2 mm

(0.079

")

0

-2 mm

(-0.079

")

15 mm

(0.591

"

)

-15 mm

(-0.591

"

)

-10 mm

(-0.394

")

0,35 m

(13.78

")

0,15 m

(5.906

"

)

0,1 m

(3.937

")

0,05 m

(1.969

")

1

1

Fig. 55: Deviation VEGAFLEX 83 in rod version in oil
1 Dead band - no measurement possible in this area

L Probe length

Repeatability

≤ ±1 mm

Variables influencing measurement accuracy
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm (0.394 in)

Additional deviation through electromag-

netic interference acc. to EN 61326

< ±10 mm (< ±0.394 in)