VEGA FIBERTRAC 31 Profibus PA User Manual
Page 73
73
11 Supplement
FIBERTRAC 31 • Profibus PA
43833-EN-131119
Ʋ ½ NPT
5 x closing cap (red) ½ NPT
Included: 3 x cable gland ½ NPT (cable: ø 6 … 12 mm),
4 x blind plug ½ NPT
Spring-loaded terminals for wire cross-section
Ʋ Massive wire, cord
0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve
0.2 … 1.5 mm² (AWG 24 … 16)
Display and adjustment module
Display element
Display with backlight
Measured value indication
Ʋ Number of digits
5
Ʋ Size of digits
W x H = 7 x 13 mm
Adjustment elements
4 keys
Protection rating
Ʋ unassembled
IP 20
Ʋ mounted into the housing without
cover
IP 40
Materials
Ʋ Housing
ABS
Ʋ Inspection window
Polyester foil
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone Ex factory
CET
Rate deviation max.
10.5 min/year
Measurement electronics temerature
Resolution
1 °C (1.8 °F)
Accuracy
±1 °C (1.8 °F)
Voltage supply
Sensor power supply
20 … 72 V DC or 20 … 253 V AC, 50/60 Hz
Profibus PA
Ʋ Non-intrinsically safe bus
9 … 32 V DC
Ʋ Intrinsically safe bus - FISCO model
9 … 17.5 V DC
Ʋ Intrinsically safe bus - ENTITY model 9 … 24 V DC
Number of sensors per DP/PA segment coupler
Ʋ Non-Ex instruments
32
Ʋ Ex instruments
10