9 Diagnostics and service

POINTRAC 31 • Profibus PA

43835-EN-131119

Code
Text mes-

sage

Cause

Rectification

F036
Faulty pro-

gram memory

– Error during software

update

– Repeat software update

– Exchanging the electronics

F037
RAM hard-

ware error

– Error in RAM

– Exchanging the electronics

F038
Slave signals

failure

– Connection cable to the

Slave instrument interrupted

– Instrument not defined as

Slave instrument

– Define instrument as Slave

– Check the connection cable

to the Slave instrument

F040
Hardware

error

– Sensor defective

– Exchanging the electronics

F041
Photomultipli-

er error

– Error in the measured value

recording

– Exchanging the electronics

F045
Error on the

current output

– Error on the current output

– Check wiring of the current

output

– Exchanging the electronics

F052
Faulty config-

uration

– Invalid parameter adjust-

ment

– Carry out a reset

F114
Error real time

clock

– Discharge accumulator

– Readjust real time clock

F122
Double ad-

dress on the

multisensor

communica-

tion bus

– Instrument addresse was

assigned several times

– Change instrument

addresses

F123
X-ray alarm

– External instruments cause

radioactive radiation

– Determine reason for X-ray

alarm

– In case of a brief X-ray

alarm, shut down the instru-

ment (switching) outputs for

this time

F124
Alarm due to

increased ra-

diation

– Radiation dose too high

– Determine reason for

increased radiation

F125
Ambient tem-

perature too

high

– Ambient temperature on

the housing outside the

specification

– Cool (heat) the instrument

or protect it with isolation

material against cold or

radiation heat