9 Diagnostics and service
POINTRAC 31 • Profibus PA
43835-EN-131119
Code
Text mes-
sage
Cause
Rectification
F036
Faulty pro-
gram memory
– Error during software
update
– Repeat software update
– Exchanging the electronics
F037
RAM hard-
ware error
– Error in RAM
– Exchanging the electronics
F038
Slave signals
failure
– Connection cable to the
Slave instrument interrupted
– Instrument not defined as
Slave instrument
– Define instrument as Slave
– Check the connection cable
to the Slave instrument
F040
Hardware
error
– Sensor defective
– Exchanging the electronics
F041
Photomultipli-
er error
– Error in the measured value
recording
– Exchanging the electronics
F045
Error on the
current output
– Error on the current output
– Check wiring of the current
output
– Exchanging the electronics
F052
Faulty config-
uration
– Invalid parameter adjust-
ment
– Carry out a reset
F114
Error real time
clock
– Discharge accumulator
– Readjust real time clock
F122
Double ad-
dress on the
multisensor
communica-
tion bus
– Instrument addresse was
assigned several times
– Change instrument
addresses
F123
X-ray alarm
– External instruments cause
radioactive radiation
– Determine reason for X-ray
alarm
– In case of a brief X-ray
alarm, shut down the instru-
ment (switching) outputs for
this time
F124
Alarm due to
increased ra-
diation
– Radiation dose too high
– Determine reason for
increased radiation
F125
Ambient tem-
perature too
high
– Ambient temperature on
the housing outside the
specification
– Cool (heat) the instrument
or protect it with isolation
material against cold or
radiation heat