beautypg.com

VEGA POINTRAC 31 Profibus PA User Manual

Page 15

background image

15

4 Mounting

POINTRAC 31 • Profibus PA

43835-EN-131119

Fig. 6: POINTRAC 31 as level detection (top view)

POINTRAC 31 lends itself well for level detection of bulk solids with

low density. Mount the instrument horizontally at the height of the

requested switching point.
Mount the source container VEGASOURCE displaced by 90° in order

to get the widest possible radiation angle.
When the sensor is covered by the medium, the radiation damping

is considerably stronger - hence, the switching point is all the more

reliable.

If the max. ambient temperature is exceeded, you must take suitable

measures to protect the instrument against overheating.
You can protect the instrument by providing a suitable insulation

against the heat or mounting the instrument further away from the

heat source.
Make sure these measures are taken into account already in the plan-

ning stage. If you want to carry out such measures later on, contact

our specialists to ensure that the accuracy of the application is not

impaired.
If these measures are not sufficient to maintain the max. ambient

temperature, you could consider using the water cooling system we

offer for POINTRAC 31.
The water cooling must also be included in the calculations for the

measuring point. Contact our specialists regarding the dimensioning

of the water cooling.

Bulk solids with low

density

Protection against heat