VEGA VEGABAR 81 Modbus and Levelmaster protocol - Operating Instructions User Manual
Page 62
62
10 Supplement
VEGABAR 81 • Modbus and Levelmaster protocol
46293-EN-131107
Ʋ External housing
– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
blind plug M20 x 1.5
or:
– 1␣x plug (depending on the version), 1␣x blind stopper
M20␣x␣1.5
Spring-loaded terminals for wire cross-
section up to
2.5 mm² (AWG 14)
Display and adjustment module
Display element
Display with backlight
Measured value indication
Ʋ Number of digits
5
Ʋ Size of digits
W x H = 7 x 13 mm
Adjustment elements
4 keys
Protection rating
Ʋ unassembled
IP 20
Ʋ mounted into the housing without
cover
IP 40
Materials
Ʋ Housing
ABS
Ʋ Inspection window
Polyester foil
Interface to the external display and adjustment unit
Data transmission
digital (I²C-Bus)
Configuration, connection cable
4-wire, screened
Cable length max.
Ʋ Sensors with signal output
4 … 20 mA, 4 … 20 mA/HART
50 m
Ʋ Sensors with signal output Profibus
PA, Foundation Fieldbus
25 m
Interface to the slave sensor
Data transmission
digital (I²C-Bus)
Configuration, connection cable
4-wire, screened
Cable length max.
25 m
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone Ex factory
CET
Rate deviation max.
10.5 min/year
Measurement electronics temerature
Resolution
1 °C (1.8 °F)