VEGA VEGABAR 83 Modbus and Levelmaster protocol - Operating Instructions User Manual
Page 49
49
8 Diagnosis, asset management and service
VEGABAR 83 • Modbus and Levelmaster protocol
46295-EN-131107
Code
Text mes-
sage
Cause
Rectification
F013
no measured
value avail-
able
– No valid measured value
available
– Gauge pressure or low
pressure, measuring cell
defective
F017
Adjustment
span too small
– Adjustment not within
specification
– Change the adjustment
according to the limit values
F025
Error in the
linearization
table
– Index markers are not con-
tinuously rising, for example
illogical value pairs
– Check linearization table
– Delete table/Create new
F036
no operable
sensor soft-
ware
– Failed or interrupted soft-
ware update
– Repeat software update
– Check electronics version
– Exchanging the electronics
– Send instrument for repair
F040
Error in the
electronics
– Hardware defect
– Exchanging the electronics
– Send instrument for repair
F041
Error in the
electronics
– No connection to the sensor
electronics
– Check connection to the
sensor electronics (with
detachable version)
F080
– General software error
– Disconnect operating volt-
age briefly
F113
Communi-
cation error,
display and
adjustment
module
Adjustment
software
PACTware out
of order
– EMC interference
– Remove EMC influences
F125
Impermissi-
ble electronics
temperature
– Temperature of the elec-
tronics in the non-specified
section
– Check ambient temperature
– Isolate electronics
– Use instrument with higher
temperature range
F260
Error in the
calibration
– Error in the calibration car-
ried out in the factory
– Error in the EEPROM
– Exchanging the electronics
– Send instrument for repair
F261
Error in the
configuration
– Error during setup
– Error when carrying out a
reset
– Repeat setup
– Repeat reset