7 Set up with the display and adjustment module
VEGAFLEX 82 • 4 … 20 mA/HART two-wire
44222-EN-130823
signal suppression, user-programmable linearization curve as well as
the measured value memory will be deleted.
Basic settings: Resetting of the parameter settings incl. special
parameters to the default values (presettings) of the respective instru-
ment. Any created false signal suppression or user-programmable lin-
earization curve as well as the measured value memory are deleted.
The following tables show the default values of the instrument. De-
pending on the instrument version or application, all menu items may
not be available or some may be differently assigned.
The menu items in bold are safety-relevant in terms of the functional
safety according to IEC 61508 (Edition 2) SIL.
Setup
Menu item
Default value
Modified value
Lock adjustment
Locked
Measurement loop name
Sensor
Units
Distance unit: mm
Temperature unit: °C
Probe length
Length of the probe Ex factory
Medium type
Bulk solid
Application
Level in the metallic vessel
Medium, dielectric constant
Granules, powder, cement /
1.5 … 3
Superimposed gas phase
Yes
Dielectric constant, upper medium (TS)
1.5
Tube inner diameter
200 mm
Max. adjustment - Level
100 %
Max. adjustment - Level
Distance: 0.000 m(d) - note block-
ing distances
Min. adjustment - Level
0 %
Min. adjustment - Level
Distance: Probe length - take dead
band into account
Accept adjustment of the level measurement? Yes
Max. adjustment - Interface
100 %
Max. adjustment - Interface
Distance: 0.000 m(d) - note block-
ing distances
Min. adjustment - Interface
0 %
Min. adjustment - Interface
Distance: Probe length - take dead
band into account
Integration time - Level
0.0 s
Integration time - Interface
0.0 s
Linearization type
Linear
Linearization - Socket correction
0 mm