11 Supplement

VEGAFLEX 86 • 4 … 20 mA/HART two-wire

41844-EN-130612

L

0

-15 mm

(-0.591

"

)

2 mm

(0.079

")

-2 mm

(-0.079

")

25 mm

(0.984

"

)

0,2 m

(7.874

")

0,15 m

(5.906

"

)

0,08 m

(3.15

")

0,1 m

(3.937

")

1

1

Fig. 53: Deviation VEGAFLEX 86 in cable version 2 mm (0.079 in), in oil
1 Dead band - no measurement possible in this area

L Probe length

L

0

-15 mm

(-0.591")

2 mm

(0.079")

-2 mm

(-0.079")

25 mm

(0.984")

0,2 m

(7.874")

0,15 m

(5.906")

0,15 m

(5.906")

0,1 m

(3.937")

1

1

Fig. 54: Deviation VEGAFLEX 86 in cable version 4 mm (0.157 in), in oil
1 Dead band - no measurement possible in this area

L Probe length

Repeatability

≤ ±1 mm

Variables influencing measurement accuracy
Specifications for the digital measured value
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm (0.394 in)

Additional deviation through electromag-

netic interference acc. to EN 61326

< ±10 mm (< ±0.394 in)

Specifications apply also to the current output

7)

7)

Also for the second current output (optional)