beautypg.com

VEGA VEGAFLEX 83 4 … 20 mA_HART - four-wire PFA insulated User Manual

Page 70

background image

70

11 Supplement

VEGAFLEX 83 • 4 … 20 mA/HART - four-wire

41835-EN-130314

0,2 m

(7.874")

0,1 m

(3.937")

0,07 m

(2.756")

1

L

0

-2 mm

(-0.079")

2 mm

(0.079")

15 mm

(0.591")

-15 mm

(-0.591")

Fig. 36: Deviation VEGAFLEX 83 in cable version (coated) in water
1 Dead band - no measurement possible in this area

L Probe length

L

2 mm

(0.079

")

0

-2 mm

(-0.079")

15 mm

(0.591

"

)

-15 mm

(-0.591")

0,35 m

(13.78")

0,2 m

(7.874")

0,2 m

(7.874")

0,05 m

(1.969")

1

1

Fig. 37: Deviation VEGAFLEX 83 in cable version (coated) in oil
1 Dead band - no measurement possible in this area

L Probe length

Repeatability

≤ ±1 mm

Variables influencing measurement accuracy
Specifications for the digital measured value
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm (0.394 in)

Additional deviation through electromag-

netic interference acc. to EN 61326

< ±10 mm (< ±0.394 in)

Specifications apply also to the current output