3 Product description

VEGABAR 86 • 4 … 20 mA/HART

45039-EN-131010

The VEGABAR 86 is a suspension pressure transmitter for level

measurement in wells, basins and open vessels. Its great flexibility

through different cable and tube versions allows the instrument to be

used in many different applications.

Measured products are liquids.
Depending on the instrument version and measurement setuo, meas-

ured products can be also viscous or contain abrasive substances.

Sensor element is the CERTEC

®

measuring cell with robust ceramic

diaphragm. The process pressure deflects the ceramic diaphragm

and causes a capacitance change in the measuring cell. This capaci-

tance change is converted into an electrical signal and outputted as

measured value via the output signal.
The CERTEC

®

measuring cell has also an integrated temperature

sensor. The temperature value is outputted through:

The display and adjustment module

The current output

The digital signal output

The measuring cell design depends on the selected pressure type.

Relative pressure: the measuring cell is open to atmosphere. The

ambient pressure is detected in the measuring cell and compensated.

It thus has no influence on the measured value.
Absolute pressure: the measuring cell is evacuated and encapsu-

lated. The ambient pressure is not compensated and does hence

influence the measured value.
Relative pressure, climate-compensated: the measuring cell is

evacuated and encapsulated. The ambient pressure is detected

through a reference sensor in the electronics and compensated. It

thus has no influence on the measured value.

The following illustration shows the installation of the CERTEC

®

measuring cell in the transmitter as well as the sealing concept.

1
2
3
4

5

Fig. 3: Front-flush installation of the CERTEC

®

measuring cell with double seal

1 Housing, transmitter

2 Measuring cell

3 Lateral seal for measuring cell

4 Additional, front seal for measuring cell

5 Diaphragm

Application area

Measured products

Measuring system

Pressure types

Seal concept