7 exchanging the electronics module, 8 software update, 9 how to proceed in case of repair – VEGA VEGABAR 86 4 … 20 mA_HART - Operating Instructions User Manual
Page 58
58
9 Diagnosis, asset management and service
VEGABAR 86 • 4 … 20 mA/HART
45039-EN-131010
The necessary serial number can be found on the type label of the
instrument or on the delivery note.
9.7 Exchanging the electronics module
In case of a defect, the user can replace the electronics module with
another one of identical type.
In Ex applications, only instruments and electronics modules with ap-
propriate Ex approval may be used.
If there is no electronics module available on site, one can be ordered
from the agency serving you.
Ordering and exchange are possible with or without sensor serial
number. The electronics module with serial number includes order-
specific data such as factory setting, seal material etc. These are not
included in the electronics module without serial number.
The serial number is stated on the type label of VEGABAR 86 or on
the delivery note.
You can find detailed information on the electronics exchange in the
booklet "Operating instructions for electronics module VEGABAR
series 80".
9.8 Software update
The following components are required to update the sensor soft-
ware:
•
Sensor
•
Voltage supply
•
Interface adapter VEGACONNECT
•
PC with PACTware
•
Current sensor software as file
You can find the actual sensor software as well as detailed infor-
mation of the procedur
"Software".
You can find information about the installation in the download file.
Caution:
Instruments with approvals can be bound to certain software ver-
sions. Therefore make sure that the approval remains effective with a
software update.
You can find detailed information on
"Approvals".
9.9 How to proceed in case of repair
You can find a repair form as well as detailed information on how to
pr
cates".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
If a repair is necessary, please proceed as follows:
•
Print and fill out one form per instrument