beautypg.com

2 principle of operation – VEGA VEGABAR 82 Save sensor - Operating Instructions User Manual

Page 8

background image

8

3 Product description

VEGABAR 82 • Slave sensor for electronic differential pressure

45050-EN-131011

This operating instructions manual applies to the following instrument

versions:

Hardware from 1.0.0

Software version from 1.0.0

The scope of delivery encompasses:

VEGABAR 82 pressure transmitter

Connection cable to the master sensor

Cable gland for the master sensor

Documentation

– this operating instructions manual

– Test certificate measuring accuracy (optional)

– Supplementary instructions manual "Plug connector for con-

tinuously measuring sensors" (optional)

– Ex-specific "Safety instructions" (with Ex versions)

– if necessary, further certificates

3.2 Principle of operation

The VEGABAR 82 slave sensor is combined with a sensor from the

VEGABAR 80 series for electronic differential pressure measurement.
The sensors are connected via a screened four-wire cable. The

measured value from the slave sensor is read in and factored into the

calculations. Parameter adjustment is carried out through the master

sensor.

The electronic differential pressure measurement is suitable for the

measurement of the following process variables:

Level

Flow

Differential pressure

Density

Interface

Fig. 2: Electronic differential pressure measurement through Master/Slave

combination

VEGABAR 82 is suitable for applications in virtually all industries. It is

used for the measurement of the following pressure types.

Gauge pressure

Scope of this operating

instructions manual

Scope of delivery

Electronic differential

pressure measurement

Measured variables

Application area