VEGA VEGABAR 82 Save sensor - Operating Instructions User Manual
Page 46
46
9 Supplement
VEGABAR 82 • Slave sensor for electronic differential pressure
45050-EN-131011
Nominal range
Overload capacity, max.
pressure
Overload capacity, min. pres-
sure
0 … 5 psi
435 psi
0 psi
0 … 15 psi
510 psi
0 psi
0 … 30 psi
725 psi
0 psi
0 … 150 psi
1300 psi
0 psi
0 … 300 psi
1900 psi
0 psi
0 … 900 psi
2900 psi
0 psi
0 … +1450 psig
(only for measuring cell ø 28 mm)
2900 psig
0 psi
Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ Temperature
+15 … +25 °C (+59 … +77 °F)
Ʋ Relative humidity
45 … 75 %
Ʋ Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Determination of characteristics
Limit point adjustment according to IEC 61298-2
Characterstic curve
Linear
Reference installation position
upright, diaphragm points downward
Influence of the installation position
< 0.2 mbar/20 Pa (0.003 psig)
Deviation (according to IEC 60770)
Specifications refer to the set span. Turn down (TD) is the ratio: nominal measuring range/set span.
Accuracy class
Non-linearity, hysteresis and re-
peatability with TD 1 : 1 up to 5 : 1
Non-linearity, hysteresis and re-
peatability with 5 : 1
0.05 %
< 0.05 %
< 0.01 % x TD
0.1 %
< 0.1 %
< 0.02 % x TD
0.2 %
< 0.2 %
< 0.04 % x TD
Influence of the product or ambient temperature
Thermal change zero signal and output span
Turn down (TD) is the relation nominal measuring range/adjusted span.
Measuring cell - Standard
Medium or ambient tem-
perature T
Accuracy class 0.05 %,
0.1 %
Accuracy class 0.2 %
Accuracy class 0.2 %
with measuring range
0.1 bar
abs
-40 … 0 °C (-40 … +32 °F) < {0.15 % + 0.015 %/K x (-
T)} x {0.5+0.5 x TD}
< {0.1 % + 0.03 %/K x (-T)}
x {0.5+0.5 x TD}
< {0.30 % + 0.03 %/K x (-
T)} x {0.5+0.5 x TD}
0 … 40 °C
(+32 … +104 °F)
< {0.0075 %/K x (T -
20 °C)} x {0.5+0.5 x TD}
< {0.005 %/K x (T - 20 °C)}
x {0.5+0.5 x TD}
< {0.015 %/K x (T - 20 °C)}
x {0.5+0.5 x TD}
40 … 100 °C
(+104 … +212 °F)
< {0.15 % } x {0.5+0.5 x
TD}
< {0.005 %/K x (T - 20 °C)}
x {0.5+0.5 x TD}
< {0.3 % } x {0.5+0.5 x TD}