6 Set up with the display and adjustment module PLICSCOM
VEGABAR 66 • Profibus PA - climate compensated
36792-EN-130321
Basic adjustment
The "Reset" "Basic adjustment" resets the values of the following
menu items:
Menu section
Function
Reset value
Basic settings
Sensor address
126
Zero/Min. adjustment
Measuring range begin
Span/Max. adjustment
Measuring range end
Density
1 kg/l
Density unit
kg/l
Damping
0 s
Linearization
Linear
Sensor-TAG
Sensor
Display
Displayed value
PA-Out
Service
Additional PA value
Secondary Value 1
Out-Scale-Unit
Volume/l
Scaling
0.00 to 100.0
Decimal point indication 8888.8
The values of the following menu items are not reset with "Reset:
Menu section
Function
Reset value
Basic settings
Unit of measurement
No reset
Temperature unit
No reset
Position correction
No reset
Display
Backlight
No reset
Service
Language
No reset
Application
No reset
Default setting
Like basic adjustment, but in addition, special parameters are reset to
default values.
9)
Peak value
The min. and max. distance values are reset to the actual value.
Additional adjustment and diagnosis options such as e.g. scaling,
simulation or trend curve presentation are shown in the following
menu schematic. You will find a detailed description of these menu
items in the operating instructions manual "Display and adjustment
module".
Optional settings
9)
Special parameters are parameters which are set customer-specifically on
the service level with the adjustment software PACTware.