6 Set up with the indicating and adjustment module

VEGAPULS 62 • Foundation Fieldbus

36506-EN-121011

Enter the requested parameters via the appropriate keys, save your

settings with [OK] and jump to the next menu item with the [ESC] and

the [->] key.

Also the vessel form can influence the measurement apart from the

medium and the application. To adapt the sensor to these measure-

ment conditions, this menu item offers you different options for vessel

bottom and ceiling in case of certain applications.

Enter the requested parameters via the appropriate keys, save your

settings with [OK] and jump to the next menu item with the [ESC] and

the [->] key.

Since a radar sensor is a distance measuring instrument, the distance

from the sensor to the product surface is measured. For indication of

the real level, an allocation of the measured distance to the percent-

age height must be carried out.
To perform the adjustment, enter the distance with full and empty ves-

sel, see the following example:

100%

0%

0,5 m (19.68

")

5 m

(196.9

")

2

1

Fig. 37: Parameter adjustment example min./max. adjustment
1 Min. level = max. meas. distance

2 Max. level = min. meas. distance

Setup/Vessel form

Setup/Adjustment