5 Setup
Electronics module • PROTRAC series 30
40104-EN-130916
5 Setup
5.1 Setup preparations
If you ordered the electronics module by stating the sensor serial
number, it is ready for the setup procedure immediately after installa-
tion and connection to power supply.
If you ordered the electronics module without stating the sensor
serial number or you are using a suitable electronics module from
stock, you first have to load the sensor data after installation.
Under the addr
The order data of the sensor is displayed after you enter the sensor
serial number.
Below the order data you will find "Sensor data for Service-DTM" as
an XML file. This file contains all sensor-specific parameters as well
as the factory settings.
Load this file with "Save target as" on your PC. Start PACTware and
under "Instrument data" go to the menu item "Additional functions",
then to "Electronics exchange". Transfer the XML file to the sensor.
5.2 Setup steps
If the instrument is used in the same application after the electronics
exchange, the previous parameter settings of the instrument must be
restored. To do this, you can use the import function of the adjustment
software PACTware with the device DTMs or the copy function of the
indicating and adjustment module.
If no copy of the parameter settings is available, you have to carry out
a fresh setup. See the operating instructions of the sensor.
With sensor serial num-
ber
Without sensor serial
number
Parameter adjustment