beautypg.com

VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART - two-wire User Manual

Page 70

background image

70

11 Supplement

VEGAPULS 61 • 4 … 20 mA/HART - two-wire

36499-EN-121011

Voltage supply
Operating voltage

Ʋ Non-Ex instrument

9.6 … 36 V DC

Ʋ Ex-ia instrument

9.6 … 30 V DC

Ʋ Ex-d-ia instrument

14 … 36 V DC

Ʋ Ex-d-ia instrument with ship approval 15 … 36 V DC

Operating voltage with illuminated indicating and adjustment module

Ʋ Non-Ex instrument

16 … 36 V DC

Ʋ Ex-ia instrument

16 … 30 V DC

Ʋ Ex-d-ia instrument

20 … 36 V DC

Interpolation protection

Integrated

Permissible residual ripple - Non-Ex, Ex-ia instrument

Ʋ for 9.6 V< U

N

< 14 V

≤ 0.7 V

eff

(16 … 400 Hz)

Ʋ for 18 V< U

N

< 36 V

≤ 1.0 V

eff

(16 … 400 Hz)

Permissible residual ripple - Ex-d-ia instrument

Ʋ for 18 V< U

N

< 36 V

≤ 1 V

eff

(16 … 400 Hz)

Load

see diagram

1200

927

500

750

250

12

18

16

9,6

14

20

15,1

22

24

26

28

30

32

34

36

V

1

4

2

3

Fig. 58: Voltage diagram
1 HART load

2 Voltage limit Ex-ia instrument

3 Voltage limit non-Ex/Ex-d instrument

4 Operating voltage

Electrical protective measures
Protection, depending on housing version

Ʋ Plastic housing

IP 66/IP 67

Ʋ Aluminium housing, stainless steel

housing - investment casting, stain-

less steel housing - electro-polished

IP 66/IP 68 (0.2 bar)

5)

Ʋ Aluminium and stainless housing, in-

vestment casting (optionally available)

IP 66/IP 68 (1 bar)

Overvoltage category

III

5)

The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.