beautypg.com

VEGA VEGABAR 67 Profibus PA User Manual

Page 57

background image

57

10 Supplement

VEGABAR 67 • Profibus PA

36743-EN-130321

Ʋ with meas. ranges from 0.4 bar

(5.8 psig)

max. 25 bar (363 psig)

14)

Pressure stage, process fitting

Ʋ Threaded fitting

316L PN 3, PVDF PN 5

15)

Ʋ Thread

316L PN 25, PVDF unpressurized

Ʋ Lock fitting

Unpressurized

Product temperature

Ʋ Suspension cable

-12 … +100 °C (+10 … +212 °F)

Ʋ Connection tube

-12 … +100 °C (+10 … +212 °F)

Vibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

16)

With version with connection tube, vibration resistance 1 g. With lengths > 0.5 m (1.64 ft) the tube

must be supported in addition.
Shock resistance

Acceleration 100 g/6 ms

17)

Electromechanical data - version IP 66/IP 67
Cable entry/plug

18)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

or:

– 2 x blind plug M20 x 1,5

Ʋ Double chamber housing

– 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5; plug M12 x 1 for the external

display and adjustment unit (optional)

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT, plug

M12 x 1 for the external display and adjustment unit

(optional)

or:

– 1 x plug (depending on the version), 1 x blind plug

M20 x 1.5; plug M12 x 1 for the external display and

adjustment unit (optional)

or:

– 2 x blind stoppers M20 x 1.5; plug M12 x 1 for the

external display and adjustment unit (optional)

Spring-loaded terminals for wire cross-

section

> 2.5 mm² (AWG 14)

14)

Limitation by the pressure-tightness of the cable connection.

15)

Limited by the overpressure resistance of the measuring cell.

16)

Tested according to the guidelines of German Lloyd, GL directive 2.

17)

Tested according to EN 60068-2-27.

18)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.