2 principle of operation – VEGA VEGABAR 55 4 … 20 mA User Manual
Page 8
8
3 Product description
VEGABAR 55 • 4 … 20 mA
36733-EN-130321
2
1
13
10
5
4
3
6
7
8
9
11
12
14
Fig. 2: Layout of the type label (example)
1 Instrument type
2 Product code
3 Approvals
4 Measuring range
5 Process pressure
6 Process temperature
7 Electronics, voltage supply
8 Protection rating
9 Order number
10 Serial number
11 Data-Matrix-Code for Smartphone-App
12 Material process seal
13 ID numbers, instrument documentation
14 Notified authority for CE marking
With the serial number, you can access the delivery data of the instru-
ment via www.vega.com, "VEGA Tools" and "serial number search".
In addition to the type label outside, you can also find the serial num-
ber on the inside of the instrument.
This operating instructions manual applies to the following instrument
versions:
•
Software from 3.82
3.2 Principle of operation
VEGABAR 55 is a pressure transmitter for use in the chemical, food
processing and pharmaceutical industries. Depending on the version,
it is used for level, gauge, absolute pressure or vacuum measure-
ment. Measured products are gases, vapours and liquids, also those
with higher temperatures.
The METEC
®
measuring cell is the measuring unit. It consists of
the ceramic-capacitive CERTEC
®
measuring cell and a special,
temperature-compensated isolating system.
The hydrostatic pressure of the product or the process pressure
causes a capacitance change in the measuring cell via the metallic
process diaphragm and the isolating liquid. This change is converted
into a corresponding output signal.
Scope of this operating
instructions manual
Application area
Functional principle