VEGA FIBERTRAC 31 4 … 20 mA_HART - four-wire User Manual
Page 41
41
6 Set up with the display and adjustment module
FIBERTRAC 31 • 4 … 20 mA/HART - four-wire
40446-EN-131119
First of all, define the function of the Slave instruments before you
define the Master instrument. The Master instrument can thus imme-
diately recognize the connected Slaves.
For this, the Slave instruments must be defined as "Summation
Slave". Select under the menu item "Setup/Application" the function
"Summation Slave".
The address setting (MGC) on the Slave instruments can be freely
selected. Only the address "0 - 0" is reserved for the Master instru-
ment.
The Master instrument must have the function "Level". For this
purpose, select under the menu item "Setup/Application" the function
"Level".
Set the address setting (MGC) on the Master instrument to "0 - 0".
You have to enter the addresses of the Slave instruments in the list of
the Master instrument. This function is not possible in the supply and
adjustment module. To do this you need PACTware with the respective
DTM.
The main menu is divided into five sections with the following func-
tions:
Setup: Settings, e.g. for measurement loop name, isotope, applica-
tion, background radiation, adjustment, signal output
Display: Settings, for example language, measured value display
Diagnosis: Information, for example, of device status, peak value,
simulation
Additional adjustments: Instrument unit, reset, date/time, copying
function
Info: Instrument name, hardware and software version, date of manu-
facture, instrument features
Check if the correct language is already set for the display. If not, you
can change the language in the menu item "Display/Language".
Start with the setup of FIBERTRAC 31.
In the main menu point "Setup", the individual submenu points
should be selected one after the other and provided with the correct
Main menu
Procedure