30 mm (1.180 in)

15 mm (0.590 in)

0

-15 mm (- 0.590 in)

- 30 mm (- 1.180 in)

1,0 m (3.280 ft)

15 m (49.21 ft)

Fig. 31: Deviation VEGAPULS 67

Influence of the ambient temperature to the sensor electronics

7

)

A

verage temperature coefficient of the zero

signal (temperature error)

0.03 %/10 K

Ambient conditions

A

mbient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

V

essel pressure

-100 … 200

kPa/-1 … 2 bar (-14.5 … 29.0 psig)

P

rocess temperature (measured on the

process fitting)

-40 … +80 °C (-40 … +176 °F)

V

ibration resistance

8)

-

P

lastic horn antenna with compression

or adapter flange

mechanical vibrations up to 2 g in the frequency
range 5 … 200 Hz

-

P

lastic horn antenna with mounting strap

mechanical vibrations up to 1 g in the frequency
range 5 … 200 Hz

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar

C

able entry/plug

9)

-

S

ingle chamber housing

l

1

x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),

1

x blind stopper M20 x 1.5

or:

l

1

x closing cap M20 x 1.5; 1 x blind stopper

M

20

x 1.5

or:

7)

R

elating to the nominal measuring range, in the temperature range

-40 … +80 °C .

8)

T

ested according to the guidelines of German Lloyd, GL directive 2.

9)

D

epending on the version M12 x 1, according to DIN 43650, Harting, 7/

8" FF.

VEGAPULS

67 • 4 … 20

mA/HART two-wire

53

10 S

upplement

32938

-EN

-100514